Abstract

This review summarises the reported clinical experience with flow diverters for the reconstruction of the parent artery of intracranial aneurysms, since their introduction 5 years ago. Over this period, the literature has documented treatment concepts and initial results. Safety concerns, some of which have proved unwarranted, have limited the use of these devices to the treatment of aneurysms which were likely to fail or had failed to be effectively treated by endosaccular coil embolisation. The emerging data now allows the risks of complications specific to this technology to be quantified and the emerging consensus on its efficacy has extended its use to include aneurysms suitable for conventional coil embolisation, in some centres. The need for antiplatelet prophylaxis will probably limit its use to anatomically complex and dissecting aneurysms in patients after spontaneous intracranial haemorrhage. This extending role highlights the need for systematic retrospective analysis of existing large case series and randomised comparative studies.

Back to top

Background to the Current use of Flow Diverter Stents

Two devices designed to induce aneurysm thrombosis by altering blood flow, commonly called flow diverters (FDs), have been undergoing evaluation for the endovascular treatment of intracranial aneurysms. The evaluation of novel endovascular implants in Stage 2 trials is an ad-hoc process which usually involves the manufacturer recruiting skilled therapists to evaluate their device in largely undefined initial populations of patients. This piecemeal accumulation of data on efficacy, safety and indications is then used to justify the device’s wider use. We feel that it is timely to take stock of FDs, and attempt in this review to analyse the published data, highlight areas of concern that remain under investigation and speculate on the future role of these devices.

The need for FDs is based on the rational that there are a proportion of intracranial aneurysms that are not effectively treated by endovascular ‘other means’ or that effective treatment using FDs is safer than alternatives. In practice the ‘other means’ are endosaccular coil embolisation (ECE), conventional high porosity stents and coils (SAC), parent artery occlusion (PAO) or neurosurgical procedures (e.g. clipping, bypass, aneurysm resections). The number of aneurysms best treated with FDs is uncertain; a proportion are due to failure of ECE caused by anatomical constraints or delayed recurrence generally ascribed to the unresolved causative haemodynamic influence. Primary ECE is successful in approximately 90% of patients selected as suitable for this treatment [1]1. Ferns SP, Sprengers MES, van Rooij WJ, Rinkel GJE, van Rijn JC, Bipat S, et al.; Coiling of Intracranial Aneurysms / Supplemental Appendix; Stroke 2009; 40:523-9 but approximately 10% will undergo retreatment because of delayed recurrence of the aneurysm [2]2. Campi A, Ramzi N, Molyneux AJ, Summers PE, Kerr RS, Sneade M, Yarnold JA, Rischmiller J, Byrne JV; Retreatment of ruptured cerebral aneurysms in patients randomized by coiling or clipping in the International Subarachnoid Aneurysm Trial (ISAT); Stroke 2007; 38:1538-1544 [PubMed]. High recurrence rates in extremely large aneurysms and aneurysms with neck widths greater than 4mm have been recognised for many years [3]3. Hope JK, Byrne JV, Molyneux AJ; Factors influencing successful angiographic occlusion of aneurysms treated by coil embolization; American Journal of Neuroradiology 1999; 20:391-399 [PubMed]. For these reasons large and giant aneurysms have been selected for treatment with FDs. A further group are unsuitable for ECE because of anatomical constraints, i.e. extremes of aneurysm size, fusiform shape, dissecting aneurysms and necks too wide to retain coils. In the past, such aneurysms have been treated by various techniques, e.g. SAC, PAO or neurosurgery.

It is inappropriate to compare the reported results of FD treatments with ECE since they have so far largely comprised aneurysms that are unsuitable for conventional endosaccular packing. Better benchmarks are alternative endovascular techniques, developed to improve the long-term stability of ECE including complex memory geometry and coated coils [4]4. White PM, Lewis SC, Gholkar A, Nahser H, Cognard C, Forrester L, et al.; Hydrogel-coated coils versus bare platinum coils for the endovascular treatment of intracranial aneurysms (HELPS): a randomised controlled trial; Lancet 2011;377:1655-1662 [5]5. Coley S, Sneade M, Clarke A, Mehta Z, Kallmes D, Cekirge S, et al.; Cerecyte Coil Trial: Procedural safety and Clinical outcomes in patients with ruptured and Unruptured Intracranial Aneurysms; Am J Neuroradiol 2012; epub Dec 2011, balloon assisted coiling [6]6. Moret J, Cognard C, Weill A, Castaings L, Rey A; The “Remodelling Technique” in the treatment of wide neck intracranial aneurysms. Angiographic results and clinical follow-up in 56 cases; Interv Neuroradiol 1997;3(1):21-35 and the use of alternative packing materials such as Onyx [7]7. Molyneux AJ, Cekirge S, Saatci I, Gál G; Cerebral Aneurysm Multicenter Swiss Onyx (CAMEO) trial: results of a prospective observational study in 20 Swiss centers; Am J Neuroradiol 2004 Jan;25(1):39-51. So far these initiatives have failed to live up to initial enthusiasm and trials have found only modest reductions in recurrence rates [4]4. White PM, Lewis SC, Gholkar A, Nahser H, Cognard C, Forrester L, et al.; Hydrogel-coated coils versus bare platinum coils for the endovascular treatment of intracranial aneurysms (HELPS): a randomised controlled trial; Lancet 2011;377:1655-1662 [5]5. Coley S, Sneade M, Clarke A, Mehta Z, Kallmes D, Cekirge S, et al.; Cerecyte Coil Trial: Procedural safety and Clinical outcomes in patients with ruptured and Unruptured Intracranial Aneurysms; Am J Neuroradiol 2012; epub Dec 2011 [8]8. Piotin M, Pistocchi S, Bartolini B, Blanc R; Intracranial aneurysm coiling with PGLA-coated coils versus bare platinum coils: long-term anatomic follow-up; Neuroradiology 2012 Apr;54(4):345-8. 2011.. Combining ECE with parent artery stenting, i.e. SAC, has several theoretical advantages (e.g. better coil retention and alterations in the blood flow patterns at the neck) but this technique has yet to demonstrate consistent reductions in recurrence rates. Moreover, they have introduced new complication risks e.g. late in-stent stenosis and delayed thromboembolic events [9]9. Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, Moret J; Stent assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms; Stroke 2010;41:110-5 [PubMed]. The literature reporting clinical and anatomical sequelae of aneurysms treated with FDs alone or with endovascular coils has emerged against this background.

Back to top

Experimental Evidence of Flow Diverters

Over the last 20 years, several researchers have tried to exploit the concept of a single endovascular stent device capable of inducing aneurysm thrombosis whilst maintaining patency of the parent artery. Systematic studies of how stent design features affect local blood flow have been conducted in experimental in-vitro studies using dyes [10]10. Lieber BB, Stancampiano AP, Wakhloo AK; Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity; Ann Biomed Eng 1997;25:460-469 [PubMed] [11]11. Imbesi SG, Kerber CW; Analysis of slipstream flow in a wide-necked basilar artery aneurysm: evaluation of potential treatment regimens; Am J Neuroradiol 2001;22(4), 721-724 [12]12. Imbesi SG, Knox K, Kerber CW; Aneurysm flow dynamics: alterations of slipstream flow for neuroendovascular treatment with liquid embolic agents; Am J Neuroradiol 2003;24(10), 2044-2049 [PubMed] and particles [13]13. Yu SC, Zhao JB; A steady flow analysis on the stented and non-stented sidewall aneurysm models; Med Eng Phys 1999;21(3), 133-141 [PubMed] [14]14. Baráth K, Cassot F, Rüfenacht DA, Fasel JH; Anatomically shaped internal carotid artery aneurysm in vitro model for flow analysis to evaluate stent effect; Am J Neuroradiol 2004; 25(10), 1750-1759 [PubMed] [15]15. Lieber BB, Livescu V, Hopkins LN, Wakhloo AK; Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow; Ann Biomed Eng 2002; 30:768 -777 [16]16. Canton G, Levy DI, Lasheras JC; Hemodynamic changes due to stent placement in bifurcating intracranial aneurysms; J Neurosurg 2005;103, 146-155 [PubMed] [17]17. Cantón G, Levy DI, Lasheras JC, Nelson PK; Flow changes caused by the sequential placement of stents across the neck of sidewall cerebral aneurysms; J Neurosurg 2005; 103(5), 891-902 [PubMed] [18]18. Seong J, Wakhloo AK, Lieber BB; In vitro evaluation of flow diverters in an elastase-induced saccular aneurysm model in rabbit; J Biomech Eng 2007;129;863-872. Particle image velocimetry has become the standard experimental technique of measuring alterations of blood flow in saccular aneurysms and the ability of FD to redirect blood flow within the parent artery [15]15. Lieber BB, Livescu V, Hopkins LN, Wakhloo AK; Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow; Ann Biomed Eng 2002; 30:768 -777 [19]19. Trager AL, Sadasivan C, Seong J, Lieber BB; Correlation between angio- graphic and particle image velocimetry quantifications of flow diverters in an in vitro model of elastase-induced rabbit aneurysms; J Biomech Eng. 2009;131:034506. These findings have been supported by in-vivo studies, which demonstrate the potential of stents to disrupt endosaccular blood flow in experimental aneurysms [20]20. Wakhloo AK, Schellhammer F, de Vries J, Haberstroh J, Schumacher M; Self-expanding and balloon-expandable stents in the treatment of carotid aneurysms: An experimental study in a canine model; AJNR Am J Neuro- radiol. 1994;15:493-502. [21]21. Turjman F, Massoud TF, Ji C, Guglielmi G, Vinuela F, Robert J; Combined stent implantation and endosaccular coil placement for treatment of experimental wide-necked aneurysms: a feasibility study in swine; Am J Neuroradiol 1994;15:1087-1090 [PubMed] [22]22. Szikora I, Guterman LR, Wells KM, Hopkins LN; Combined use of stents and coils to treat experimental wide necked carotid aneurysms: preliminary results; Am J Neuroradiol 1994; 15:1091-1102 [PubMed] [23]23. Massoud TF, Turjman F, Ji C, Viñuela F, Guglielmi G, Gobin YP, Duckwiler GR; Endovascular treatment of fusiform aneurysms with stents and coils: technical feasibility in a swine model; Am J Neuroradiol 1995; 16: 1953-1963 [24]24. Geremia G, Haklin M, Brennecke L; Embolization of experimentally created aneurysms with intravascular stent devices; Am J Neuroradiol 1994;15:1223-1231 [PubMed] [25]25. Wakhloo AK, Tio FO, Lieber BB, Schellhammer F, Graf M, Hopkins LN; Self-expanding Nitinol stents in canine vertebral arteries: Hemodynamics and tissue response; Am J Neuroradiol 1995;16:1043-1051 [PubMed].  Using a rabbit model, Kallmes et al. showed that the orifices of side branches remain patent despite general endothelisation of implanted FDs [26]26. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ; A new endoluminal, flow-disrupting device for treatment of saccular aneurysms; Stroke. 2007;38:2346 -2352 [27]27. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ; A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms; AJNR Am J Neuroradiol. 2009;30:1153-1158 [PubMed].

Quantification of endosaccular flow on digital subtraction angiography can be performed using time-density curves [28]28. Tenjin H, Asakura F, Nakahara Y; Evaluation of intraaneurysmal blood velocity by time-density curve analysis and digital subtraction angiography; Am J Neuroradiol 1998;19:1303-1307 [PubMed] or washout models [29]29. Kawanishi M, Nagasawa S, Ohta T, Kajimoto S, Kondoh S; Simulation study on therapeutic vertebral artery occlusion for VA-PICA giant aneurysm; Neurol Res 1994;16, 100-103 [PubMed] and together with computational fluid dynamic simulation techniques [30]30. Kim M, Taulbee DB, Tremmel M, Meng H; Comparison of two stents in modifying cerebral aneurysm hemodynamics; Ann Biomed Eng. 2008;36: 726 -741 have been used to study the haemodynamic effects of stenting [31]31. Asakura F, Tenjin H, Sugawa N, Kimura S, Oki F; Evaluation of intraaneurysmal blood flow by digital subtraction angiography: blood flow change after coil embolization; Surg Neurol 2003; 59(4), 310-319. [32]32. Rudin S, Wang Z, Kyprianou I, Hoffmann KR, Wu Y, Meng H, Guterman LR, Nemes B, Bednarek DR, Dmochowski J, Hopkins LN; Measurement of flow modification in phantom aneurysm model: comparison of coils and a longitudinally and axially asymmetric stent - initial findings; Radiology 2004;231(1), 272-276. [PubMed]. These data have been used to inform the design of flow diverters [33]33. Sadasivan C, Lieber BB, Gounis MJ, Lopes DK, Hopkins LN; Angiographic quantification of contrast medium washout from cerebral aneurysms after stent placement; Am J Neuroradiol 2002;23(7):1214-1221. [PubMed] [34]34. Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB; An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model; Stroke 2009; 40(3):952-958 [PubMed]. Such work has established porosity (defined as the proportion of open area to total area of the stent) and pore density (defined as number of pores per area) as important parameters and a porosity threshold of around 70% as the optimum compromise between stent flexibility and efficient flow diversion [18]18. Seong J, Wakhloo AK, Lieber BB; In vitro evaluation of flow diverters in an elastase-induced saccular aneurysm model in rabbit; J Biomech Eng 2007;129;863-872 [35]35. Liou TM, Li YC, Juan WC; Numerical and experimental studies on pulsatile flow in aneurysms arising laterally from a curved parent vessel at various angles; J Biomech 2007;40(6), 1268-75. Other important design features are filament size [15]15. Lieber BB, Livescu V, Hopkins LN, Wakhloo AK; Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow; Ann Biomed Eng 2002; 30:768 -777 and net design, which influence the implant’s effect on side branches. In vitro studies have shown that despite substantial coverage of branch arteries, flow reductions in side branches are modest [18]18. Seong J, Wakhloo AK, Lieber BB; In vitro evaluation of flow diverters in an elastase-induced saccular aneurysm model in rabbit; J Biomech Eng 2007;129;863-872 [36]36. Appanaboyina S, Lohner R, Scrivano E, Miranda C, Lylyk P, Putman C, et al.; Computational modelling of blood flow in side arterial branches after stenting of cerebral aneurysm; Int J Comput Fluid Dynamics 2008;22:669-676. However, it is worth remembering that the effective porosity of deployed FD stents depends on other factors such as correct sizing, degrees of opening [37]37. Aurboonyawat T, Blanc R, Schmidt P, et al.; An in vitro study of Silk stent morphology; Neuroradiology 2011;53:659-67. [PubMed] and position in relation to the geometry of the parent vessel and aneurysm neck.

Back to top

Stent Assisted Coiling

The principle goal of all aneurysm treatments is exclusion of the aneurysm from the circulation; placing a stent across the aneurysm neck to obstruct blood flow into the sac achieves this objective. Covered stents with zero porosity have been used in cerebral vessels but are generally too stiff for intracranial navigation [38]38. Pero G, Denegri F, Valvassori L, Boccardi E, Scialfa G; Treatment of a middle cerebral artery giant aneurysm using a covered stent: Case report; J Neurosurg 2006;104(6):965-8 [PubMed]. Porous stents were used first to treat extra-cranial aneurysms [39]39. Mase M, Banno T, Yamada K, Katano H; Endovascular stent placement for multiple aneurysms of the extracranial internal carotid artery: Technical case report; Neurosurgery 1995;37:832-835. [PubMed] [40]40. Hurst R, Haskal Z, Zager E; Endovascular stent treatment of cervical internal carotid artery aneurysms with parent vessel preservation; Surg Neurol 1998;50:313-317 [41]41. Horowitz M, Miller G, Meyer Y, Carstens G, Purdy P; Use of intravascular stents in the treatment of internal carotid extracranial vertebral artery pseudoaneurysms; Am J Neuroradiol 1996;17:693-6 [PubMed] and early reports of their use in intracranial arteries involved aneurysms associated with stenotic arterial disease [42]42. Marks MP, Dake MD, Steinberg GK, Norbash AM, Lane B; Stent placement for arterial and venous cerebrovascular disease: Preliminary experience; Radiology 1994. [PubMed] or fusiform shapes [43]43. Higashida RT, Smith W, Gress D, Urwin R, Dowd CF, Balousek PA, Halbach VV; Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery. Case report and review of the literature; J Neurosurg 1997;87(6):944-9 [PubMed] [44]44. Brassel F, Rademaker J, Haupt C, Becker H; Intravascular stent placement for a fusiform aneurysm of the posterior cerebral artery: Case report; Eur Radiol 2001;11:1250-1253 [PubMed].  With the availability of more flexible but high porosity designs, intracranial deployment became possible and several practitioners reported treatments with a combination of stents and coils [45]45. Lanzino G, Wakhloo AK, Fessler RD, Hartney ML, Guterman LR, Hopkins LN; Efficacy and current limitations of intravascular stents for intracranial internal carotid, vertebral, and basilar artery aneurysms; J Neurosurg 1999;91(4):538-546 [PubMed] [46]46. Mericle RA, Lanzino G, Wakhloo AK, Guterman LR, Hopkins LN; Stenting and secondary coiling of intracranial internal carotid artery aneurysm: Technical case report; Neurosurgery 1998;43:1229-1234 [PubMed] [47]47. Lylyk P, Cohen JE, Ceratto R, Ferrario A, Miranda C; Combined endovascular treatment of dissecting vertebral artery aneurysms by using stents and coils; J Neurosurg 2001; 94:427-432 [PubMed] [48]48. Zenteno M, Modenesi Freitas JM, Aburto-Murrieta Y, Koppe G, Machado E, Lee A; Balloon-expandable stenting with and without coiling for wide-neck and complex aneurysms; Surg Neurol 2006; 66(6), 603-610 [PubMed] [49]49. Wakhloo AK, Lanzino G, Lieber BB, Hopkins LN; Stents for intracranial aneurysms: the beginning of a new endovascular era? Neurosurgery 1998;43:377-379 [PubMed]. There have been reports of successful sole stenting (i.e. without adjuvant endosaccular coils) using conventional high porosity stents [50]50. Yang X, Wu Z, Mu S, Li Y, Lv M; Endovascular treatment of giant and large intracranial aneurysms using the neuroform stent-assisted coil placement; Neurol Res 2008;30: 598-602 [PubMed], [51]51. Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, Moret J; Stent assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms; Stroke 2010;41:110-5 [PubMed] but the long-term anatomical results have been inconsistent [52]52. Lylyk P, Cohen JE, Ceratto R, Ferrario A, Miranda C; Endovascular reconstruction of intracranial arteries by stent placement and combined techniques; J Neurosurg 2002; 97:1306-1313 [PubMed] [53]53. Szikora I, Berentei Z, Kulcsar Z, Barath K, Berez A, Bose A, Nyary I; Endovascular treatment of intracranial aneurysms with parent vessel reconstruction using balloon and self expandable stents; Acta Neurochir 2006; 148:711-723 [54]54. Ahn JY, Han IB, Kim TG, Yoon PH, Lee YJ, Lee BH, Seo SH, Kim DI, Hong CK, Joo JY; Endovascular treatment of intracranial vertebral artery dissections with stent placement or stent-assisted coiling; Am J Neuroradiol 2006;27(7):1514-2 [55]55. Zenteno MA, Santos-Franco JA, Freitas-Modenesi JM, Gómez C, Murillo-Bonilla L, Aburto-Murrieta Y, Díaz-Romero R, Nathal E, Gómez-Llata S, Lee A; Use of the sole stenting technique for the management of aneurysms in the posterior circulation in a prospective series of 20 patients; J Neurosurg. 2008;108(6):1104-18. These include attempts to decrease the effective porosity of the stent by using double or overlapping stents [53]53. Szikora I, Berentei Z, Kulcsar Z, Barath K, Berez A, Bose A, Nyary I; Endovascular treatment of intracranial aneurysms with parent vessel reconstruction using balloon and self expandable stents; Acta Neurochir 2006; 148:711-723 [56]56. Benndorf G, Campi A, Schneider GH, Wellnhofer E, Unterberg A; Overlapping stents for treatment of a dissecting carotid artery aneurysm; J Endovasc Ther 2001;8(6):566-70 [PubMed] [57]57. Benndorf G, Herbon U, Sollmann WP, Campi A; Treatment of a ruptured dissecting vertebral artery aneurysm with double stent placement: Case report; Am J Neuroradiol 2001; 22: 1844-1848 [PubMed] [58]58. Islak C, Kocer N, Albayram S; Bare stent-graft technique: a method of endoluminal vascular reconstruction for the treatment of giant and fusiform aneurysms; Am J Neuroradiol 2002;23:1589-1595 [PubMed] [59]59. Doerfler A, Wanke I, Egelhof T, Stolke D, Forsting M; Double-stent method: therapeutic alternative for small wide-necked aneurysms. Technical note; JNeurosurg 2004; 100(1):150-4 [PubMed]. To date no comparative trial of these techniques has been reported.

There have been several single centre reports of SAC intracranial aneurysm treatments [60]60. Fiorella D, Albuquerque FT, Woo H, Rasmussen PA, Masaryk TJ, McDougall CG; Neuroform stent assisted aneurysm treatment: evolving treatment strategies, complications and results of long term follow-up; J NeuroIntervent Surg 2010; 2:16-22 [50]50. Yang X, Wu Z, Mu S, Li Y, Lv M; Endovascular treatment of giant and large intracranial aneurysms using the neuroform stent-assisted coil placement; Neurol Res 2008;30: 598-602 [PubMed] [9]9. Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, Moret J; Stent assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms; Stroke 2010;41:110-5 [PubMed] [61]61. Lylyk P, Ferrario A, Pasbón B, Miranda C, Doroszuk G; Buenos Aires experience with the Neuroform self-expanding stent for the treatment of intracranial aneurysms; J Neurosurg 2005; 102(2):235-41 and one multicentre registry [62]62. Mocco J, Snyder KV, Albuquerque FC, Bendok BR, Alan SB, Carpenter JS, et al.; Treatment of intracranial aneurysms with the Enterprise stent: a multicenter registry; J Neurosurg 2009;110:35-9. The larger series of Fiorella et al. [60]60. Fiorella D, Albuquerque FT, Woo H, Rasmussen PA, Masaryk TJ, McDougall CG; Neuroform stent assisted aneurysm treatment: evolving treatment strategies, complications and results of long term follow-up; J NeuroIntervent Surg 2010; 2:16-22 and Lylyk et al. [61]61. Lylyk P, Ferrario A, Pasbón B, Miranda C, Doroszuk G; Buenos Aires experience with the Neuroform self-expanding stent for the treatment of intracranial aneurysms; J Neurosurg 2005; 102(2):235-41 reported morbidity rates of 5.3% and 7.4% and mortality of 2.8% and 4.6% respectively. In the report of a multicentre registry concerning 141 patients, the morbidity was 2.8% and mortality 2% [62]62. Mocco J, Snyder KV, Albuquerque FC, Bendok BR, Alan SB, Carpenter JS, et al.; Treatment of intracranial aneurysms with the Enterprise stent: a multicenter registry; J Neurosurg 2009;110:35-9. Indications for the use of SAC in early reports were generally similar to those currently applied for FDs but since FDs have been available, stents such Neuroform  (Stryker) or Enterprise (Codman), are more often used for other indications. They still represent the endovascular treatment closest to arterial reconstruction with FDs and are therefore the best historic comparator of complication rates and safety.

Aneurysm recurrence rates after SAC treatments have been reported as lower than ECE but, as described above, the reduction has been modest. Sedat et al. [63]63. Sedat J, Chau Y, Mondot L, Vargas J, Szapiro J, Lonjon M; Endovascular occlusion of intracranial wide-necked aneurysms with stenting (Neuroform) and coiling: mid-term and long-term results; Neuroradiology 2009;51:401-9 [PubMed] reported recurrence on first follow up angiograms in 9.5% of aneurysms in a small series of aneurysms with mean maximum dimensions of 11.3mm and mean neck sizes of 5.33mm. In a larger series, which included dissecting and fusiform aneurysms, Fiorella et al. [64]64. Fiorella D, Albuquerque FC, Deshmukh VR, McDougall CG; Usefulness of the Neuroform stent for the treatment of cerebral aneurysms: results at initial (3-6-mo) follow-up; Neurosurgery 2005; 56:1191-201 reported recurrences in 22% which was similar to the residual aneurysm filling reported by Biondi et al., who had an 11% retreatment rate after Neuroform (Stryker) assisted coiling [65]65. Biondi A, Janardhan V, Katz JM, Salvaggio K, Riina HA, Gobin YP; Neuroform stent-assisted coil embolization of wide-neck intracranial aneurysms: strategies in stent deployment and midterm follow-up; Neurosurgery 2007;61(3):460-8; discussion 468-9. In a large single series comparing treatments with and without adjuvant stents, Piotin et al. [9]9. Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, Moret J; Stent assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms; Stroke 2010;41:110-5 [PubMed] found that recurrence rates were reduced (14.9% SAC versus 33% ECE) and this improvement applied for both small and larger aneurysm groups. However, the use of stents increased procedure related morbidity (7.4% SAC versus 3.8% ECE) and mortality (4.6% SAC versus 1.2% ECE) rates. The HELPS trial also found that SAC increased the risk of untoward events compared to ECE alone [66]66. White PM, Lewis SC, Nahser H, Sellar RJ, Goddard T, Gholkar A; Hydrocoil Endovascular Aneurysm Occlusion and Packing Study (HELPS Trial): Procedural Safety and Operator-Assessed Efficacy Results; Am J Neuroradiol 2008;29(2):217-23. [PubMed].

Back to top

Current Practice

Since 2007, two FDs produced with the aim of uncoupling blood flow between parent artery and aneurysm sac have been available in Europe. They have a woven design of high wire density and low porosity. The treatment goals are: (a) to reconstruct the parent vessel and (b) to redirect blood flow along the longitudinal axis of the parent vessel thereby modifying the haemodynamic forces acting on the aneurysm [67]67. D’Urso PI, Lanzino G, Cloft HJ, Kallmes DF; Flow Diversion for Intracranial Aneurysms; Stroke 2011; 42: 2363-8. Once deployed, the FD creates a physical barrier across the aneurysm neck disrupting blood flow in and out of the sac. The barrier introduced by the FD is subsequently augmented by the growth of endothelium and a neo-intima on its inner surface, consolidating the uncoupling. Endosaccular stagnation and thrombosis of blood inside the sac may take several weeks to develop. It is best demonstrated by MR or CT scanning. Once achieved, endothelialisation of the FD completes the reconstruction of the parent vessel - aneurysm complex [68]68. Kamran M, Yarnold J, Grunwald IQ, Byrne JV; Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema; Neuroradiology 2011; 53: 501-8 [PubMed].

Instructions of use for both available FDs are broadly similar; the major difference is that adjuvant endosaccular coils are recommended for the SILK flow diverter (SFD) (Balt Extrusion, Montmorency, France). The SFD received Swiss Community approval (i.e., CE marking) for use in January 2008. The Pipeline embolisation device (PED) (Covidien/ev3, Irvine, California) also has a CE mark and was approved by the FDA (Food and Drug Administration, US) in 2011 for the treatment of infra- and supraclinoid internal carotid artery aneurysms without ECE.

The SFD is a flexible, self-expanding, braided mesh stent with flared ends and a pore size of 110 – 250 μm. It is constructed from 48 braided nitinol (nickel-titanium alloy) and 4 platinum microfilaments ~35μm, which provides 35 – 55% metal coverage when expanded to its nominal diameter. Currently, SFDs are available in a range of diameters (3 – 5.5mm) and lengths (15 – 40 mm). The platinum strands act as radiopaque markers for visualisation under fluoroscopy. Second and third generation devices have been developed with lower porosity and increased the radial force. A SFD (termed SILK-plus) is now available with 8 platinum markers and with the option of tapered diameters to improve the conformity of the device to arteries of varying diameters. The PED is similar and is constructed from 48 microfilaments composed of platinum (25%) and nickel-cobalt-chromium alloy (75%). It is available in sizes similar to SFD (diameters 2.5 – 5 mm, lengths 10 – 35 mm). It is designed to provide 30 – 35% metal coverage of the target vessel’s inner surface with a pore size of 0.02 – 0.05 mm2 at nominal diameter. The PED is attached to a pusher wire, compatible with a 3F (0.027” ID) microcatheter. The pusher wire has a platinum coil tip, which extends 15 mm beyond the distal edge of the PED from which it is detached once deployed.

The devices are delivered via microcatheter (SFD using a 0.023”ID and PED a 0.027”ID microcatheter) requiring 6F guide catheter support. A tri-axial approach, including a 6F, 90 cm long sheath and a 6F distal access catheter is commonly used to provide better support and stability during deployment. Deployment involves positioning the FD’s tip distal to the target aneurysm. This is followed by pushing the device, applied to a delivery microwire, to the tip of the delivery microcatheter. The system is then aligned with the selected ‘landing position’ distal to the aneurysm orifice under x-ray fluoroscopy. The SFD is deployed by unsheathing it from the microcatheter by a combination manoeuvre involving pushing the delivery wire and retrieving the microcatheter to encourage the SFD to expand. The SFD can be retrieved into the microcatheter and removed or repositioned if less than 90% of its length has been extruded from the microcatheter. No retrieval is possible thereafter. For deployment of the PED, the delivery wire is held while the distal 1/3 of the PED is carefully unsheathed.  Once the unsheathed segment begins to expand, its distal end is released from the delivery wire either spontaneously or by clockwise rotation of the wire. The proximal segment of the PED is then deployed by gentle advancement of the wire. The PED is not retrievable, but at any point up to final deployment it may be captured and removed from the body [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47.

Endosaccular coils can be placed before the FD is deployed or afterwards via a second microcatheter positioned prior to SFD deployment, a technique often referred to as ‘jailing’ [70]70. Lubicz B, Collignon L, Raphaeli G, Pruvo J-P, Bruneau M, De Witte O, et al.; Flow-Diverter Stent for the Endovascular Treatment of Intracranial Aneurysms; Stroke 2010; 41: 2247-53 [PubMed]. Multiple overlapping FDs may be needed to cover wide necked or large fusiform aneurysms. In this case, after the first FD is fully deployed, the microcatheter may be re-advanced over the indwelling delivery wire, which is then exchanged for a second FD. The FDs are sometimes deployed within a conventional higher radial force stent and expanding a balloon within the deployed FD (i.e. stentplasty) can help to ensure that the device is fully expanded and conforms well to the vessel wall. Recently the Surpass device (Surpass, Tel Aviv, Israel) also obtained CE approval (2010) and has now started clinical trials in Europe.

Back to top

Prophylaxis against Thromboembolism and Parent Artery Thrombosis

Stent implants are likely to cause local vessel thrombosis and thromboembolism and a regimen of prophylaxis with antiplatelet drugs (APDs) is generally recommended.  Data of this risk have been translated from stent use in cardiology and interventional radiology and their use for intracranial aneurysm treatment has not been tested in specific randomised trials. In practice, protocols have developed which balance the need to prevent thromboembolic complications against an increased risk of iatrogenic bleeding either intra-cranial or extra-cranial. The risk of the latter is illustrated by reports of serious non-neurological bleeding in up to 4.5% of FD treatments attributed to APDs. APD prophylaxis is best given before the FD is deployed; however, after recent aneurysm rupture there is a natural reluctance to give long acting APDs until the ruptured aneurysm has been secured.

Various pre-treatment regimes have been reported for non-emergency treatments; these vary in length from hours to days. The prescriptions usually involve a combination of aspirin and a thienopyridine (e.g. clopidogrel, ticlopidine or prasugrel). Clopidogrel is a prodrug and activated in the liver by cytochrome P450 enzymes, including CYP2C19. Platelet inhibition can be demonstrated 2h after a single dose of oral clopidogrel, but the onset of action may be slower so a loading dose is usually administered a few hours or days prior to FD placement [71]71. Byrne JV; Tutorials in Endovascular Neurosurgery and Interventional Neuroradiology; Springer 2012; Tutorial 6: 97-107. The antiplatelet effect of aspirin is faster but given orally its effect increases over time and therefore both drugs are usually prescribed together. Examples of reported pre-treatment prescription protocols are 75 – 300mg aspirin and 75mg of clopidogrel for 3 – 7 days or a short regime of 300mg of both drugs 1 – 2 days prior to FD placement, though a loading dose of 600mg of clopidogrel has been shown to be better than 300mg in coronary arteries [72]72. Mangiacapra F, Muller O, Ntalianis A, Trana C, et al.; Comparison of 600 versus 300-mg Clopidogrel loading dose in patients with ST-segment elevation myocardial infarction undergoing primary coronary angioplasty; Am J Cardiol. 2010;106:1208-11 [PubMed].

If FDs are used without pre-treatment, e.g. after recent subarachnoid haemorrhage, 500mg of aspirin given intravenously is effective in minutes and can be given with oral loading doses of a thienopyridine. Additional anticoagulation with heparin is used during the deployment procedure but normally not continued afterwards. Some centres also prescribe nimodipine to prevent vessel spasm during FD delivery [73]73. Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al.; Flow-Diverter Silk Stent for the Treatment of Intracranial Aneurysms: 1-year Follow-Up in a Multicenter Study; Am J Neuroradiol. Epub 2012 Feb 2.

Antiplatelet drugs should be given post-treatment for 3 – 6 months. A typical maintenance regime is 75mg – 300mg of aspirin and 75mg clopidogrel per day for 1 – 6 months though exact dosages vary depending on formulations in different countries. Many centres stop clopidogrel before aspirin, e.g. after 3 months, whilst maintaining aspirin for 6 months. Because of individual variations in the effective dose of both aspirin and clopidogrel, testing the patient’s platelet activity after giving loading doses is an important precaution. About 15% of patients don’t respond to oral clopidogrel [71]71. Byrne JV; Tutorials in Endovascular Neurosurgery and Interventional Neuroradiology; Springer 2012; Tutorial 6: 97-107. A variety of in vitro platelet activity measurement methods are available. Their sensitivity is influenced by several factors, including the length of time since APDs were first administered. The validation of most tests is controversial which contributes to considerable confusion regarding their role in FD treatment.

Back to top

Selection of Aneurysms for Treatment with Flow Diverters

The types of aneurysms that have been selected for treatment with FDs are predominantly those considered at high risk of recurrence after ECE, those with complex shapes or aneurysms that have recurred after other treatments. This is largely pragmatic and in theory most saccular aneurysms could be treated. No clear consensus on indications has yet emerged. Reconstruction of the parent artery with FDs is the most elegant solution for all aneurysms, but in practice concerns remain about deploying FDs at arterial bifurcations because of the risk of branch artery occlusion, and for small aneurysms where recurrence after ECE is uncommon. Other outstanding questions are a possible higher rate of complications after treatments of fusiform aneurysms, particularly those in the posterior circulation and the uncertain benefit of adjuvant endosaccular coils. There have been few reports of patients treated after acute SAH [74]74. Matouk C, O'Kelly C, Blis M, Sarma D, Spears J, et al.; Pipeline embolisation device reconstruction of ruptured intracranial aneurysms: Report of 2 cases; Proceedings of the 45th. Annual Congress of the Canadian Neurological Sciences Federation 2010 Quebec. Canadian Journal of Neurological Sciences 2010;3 (supplement 1): 888-889 [75]75. McAuliffe W, Wenderoth JD; Immediate and midterm results following treatment of recently ruptured intracranial aneurysms with the Pipeline embolization device; AJNR Am J Neuroradiol. 2012 Mar;33(3):487-93 and too few for analysis, though SAC is increasingly used in this situation and numbers are likely to increase.

Back to top

Literature on the use of Pipeline and Silk FDs in practice

There have been several clinical reports of the use of PED; 2 trials [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed] [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads], case series [78]78. Fiorella D, Woo HH, Albuquerque FC, Nelson PK; Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device; Neurosurgery 2008;62(5):1115-20.  [79]79. Fiorella D, Kelly ME, Albuquerque FC, Nelson PK; Curative reconstruction of a giant mid basilar trunk aneurysm with the pipeline embolization device; Neurosurgery 2009; 64:212-7 [80]80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42  [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47, case reports [81]81. Cruz J P, Chow M, O'Kelly C, Marotta B, Spears J, Montanera W, Fiorella D, Marotta T; Delayed Ipsilateral Parenchymal Haemorrhage following Flow Diversion for the Treatment of Anterior Circulation Aneurysms; AJNR 2012; 33:603-8 [82]82. Fiorella D, Hsu D, Woo HH, Tarr RW, Nelson PK; Very late thrombosis of a pipeline embolization device construct: case report; Neurosurgery 2010; 67: ons E313-ons E314 [83]83. Klisch J, Turk A, Turner R, Woo HH, Fiorella D; Very late thrombosis of flow-diverting constructs after the treatment of large fusiform posterior circulation aneurysms; Am J Neuroradiology 2011; 32(4):627-32 [84]84. van Rooij WJ, Sluzewski M; Perforator infarction after placement of a pipeline flow-diverting stent for an unruptured A1 aneurysm; Am J Neuroradiol 2010;31:43-44 [85]85. Hampton T, Walsh D, Tolias C, Fiorella D; Mural destabilization after aneurysm treatment with a flow-diverting device: A report of two cases; J NeuroInterv Surgery 2011;3(2):167-171 [86]86. Fiorella D, Albuquerque F, Gonzalez F, McDougal CG, Nelson PK; Reconstruction of the right anterior circulation with the Pipeline embolization device to achieve treatment of a progressively symptomatic, large carotid aneurysm; J NeuroInterv Surgery 2009; 2:31-37 and abstracts of conference presentations [87]87. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Gubucz I, Nelson PK, et al.; Effect of flow modification on aneurysm induced mass effect; Proceedings of the 19th Symposium Neuroradiologicum - the World Congress of Diagnostic and Therapeutic Neuroradiology; Bologna, Italy. Neuroradiology 2010;23(1):324 [88]88. Phillips T, Mitchell P, Dowling R, Yan B; Endovascular treatment of intracranial aneurysms with new generation flow diverting stents. Early experience in an Australian neurointerventional centre; Proceedings of the 61st annual scientific meeting of the Royal Australian and New Zealand College of Radiologists (RANZCR) 2010; Oct 14-17; Perth, Australia Journal of Medical Imaging Radiation & Oncology 2010 Oct; 54(Supplement1):A122 [89]89. Hartmann M, Rohde S, Braun C, Hahnel S, Bendszus M; Endovascular treatment of cerebral aneurysms with the pipeline embolization device; Proceedings of the Jahrestagung der Deutschen Gesellschaft fur Neuroradiologie (DGNR) 2010; Sept 22-25; Mannheim Germany. Clinical Neuroradiology 2010;20(3):190-1 [90]90. O'Kelly C, Spears J, Chow M, Wong J, Silvaggio J, Boulton M, et al.; Canadian experience with the pipeline embolization device for repair of unruptured intracranial aneurysms; Canadian Journal of Neurological Sciences Conference: 46th Annual Congress of the Canadian Neurological Sciences Federation, 2011; 38(3) supp 1:S31 [91]91. Sararols L, Castillo L, Graell X, Macho J, San-Roman L, Macaya J, et al.; Right giant internal carotid artery bifurcation aneurism: Presentation with homonymous left hemianopsia and successful treatment with intraneurismatic bypass; Neuro-Ophthalmology Conference: 10th. Swiss Neuro-Ophthalmology Society, EUNOS Meeting Barcelona Spain. Conference Publication 2011; 35: S65. These involve some duplication of patients in different reports, e.g. 6 patients from Lylyk et al. 2009 [80]80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42 and 9 patients from Szikora et al. 2010 [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47, were included in the PITA trial [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed]. Though all these studies make valid points, the two trials provided much of the data for this review. These trials, called Pipeline Intracranial Treatment of Aneurysms (PITA) [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed] and Pipeline for Uncoilable or Failed aneurysmS (PUFS) [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads] were prospective single arm multicentre trials. The PITA collected data on 31 patients treated in 4 centres for 31 aneurysms selected as unsuitable for ECE. The PUFS study has not been published in a peer review journal but data submitted to the U.S. Food and Drug Administration in 2011 are available [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads]. These report details of 108 patients and 110 aneurysms treated in 10 centres. The published data on the SFD comprises a multicentre registry [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central], case series [93]93. Lubicz B, Collignon L, Raphaeli G, et al.; Flow-diverter stent for the endovascular treatment of intracranial aneurysms: a prospective study in 29 patients with 34 aneurysms; Stroke 2010;41:2247-53 [PubMed] [94]94. Kulcsár Z, Ernemann U, Wetzel SG, et al.; High-profile flow diverter (Silk) implantation in the basilar artery: efficacy in the treatment of aneurysms and the role of the perforators; Stroke 2010;41:1690-96 [73]73. Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al.; Flow-Diverter Silk Stent for the Treatment of Intracranial Aneurysms: 1-year Follow-Up in a Multicenter Study; Am J Neuroradiol. Epub 2012 Feb 2 [95]95. Wagner A, Cortsen M, Hauerberg J, Romner B, Wagner MP; Treatment of intracranial aneurysms: Reconstruction of the parent artery with flow-diverting (Silk) stent; Neuroradiology 2011; Epub Sept 2011 [96]96. Cirillo L, Dalla'olio M, Princiotta C, Simonetti L, Stafa A, Toni F, Leonardi M; The use of flow-diverting stents in the treatment of giant cerebral aneurysms: Preliminary results; The Neuroradiology Journal 2010; 23: 220-224 [97]97. Kulcsár Z, Wetzel SG, Augsburger L, Gruber A, Wanke I, Rüfenacht DA; Effect of flow diversion treatment on very small ruptured aneurysms; Neurosurgery 2010; 67: 789-93 [PubMed] and case reports [98]98. Turowski B, Macht S, Kulcsar Z, et al.; Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (Silk stent): do we need to rethink our concepts? Neuroradiology 2011;53:37- 41. [99]99. Pumar JM, Garcia-Dorrego R, Nieto A, Vazquez-Herrero F, Blanco-Ulla M, Vazquez-Martin A; Vascular reconstruction of a fusiform basilar aneurysm with the Silk embolization system; J Neurointerv Surg 2010; 2: 242-4 [PubMed] [100]100. Leonardi M, Dall'olio M, Princiotta C, Simonetti L; Treatment of carotid siphon aneurysms with a microcell stent: A case report; Interv Neuroradiol 2008;14(4);429-434 [101]101. Rasskazoff S, Silvaggio J, Brouwer PA, Kaufmann A, Nistor A, Iancu D; Endovascular treatment of a ruptured blood blister-like aneurysm with a flow diverting stent; Interv Neuroradiol 2010;16(3):255-258. [PubMed] [PubMed Central].

Back to top

Reports of Procedures

Descriptions of procedures cover 3 areas; clinical adverse event (CAE), technical adverse event (TAE) and the immediate effect of FDs on flow in aneurysm and parent artery. Published reports only record a minority of treatments because more than 1500 aneurysms had been treated by March 2010 with SFDs but the proportion of these procedures reported in the literature is less that 20%.  Byrne et al. [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central] conducted the first international multicentre registry (SFD Registry) and Berge J et al. [73]73. Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al.; Flow-Diverter Silk Stent for the Treatment of Intracranial Aneurysms: 1-year Follow-Up in a Multicenter Study; Am J Neuroradiol. Epub 2012 Feb 2 a multicentre retrospective review of patients treated in 6 French centres. Both studies reported data from consecutive treatments over a defined time period. In the SFD Registry, data on 70 patients treated at 18 participating centres worldwide were evaluated for the device’s technical performance and safety. In the French study 68 patients were treated for 77 unruptured complex or recurrent aneurysms and followed up at 6 and 12 months. In both studies, the majority of procedures were performed using SFD alone (86% and 78% respectively) and the cohorts contained a higher proportion of fusiform or dissecting aneurysms (37% and 32% respectively) than the PITA. Adjuvant coils were used in a small proportion of aneurysms (14% and 21% respectively). In PITA a single PED was used in 58% of aneurysms, and the aneurysm was covered by multiple PEDs in 42%. Additional coils were used in 52%. In PUFS the aneurysm location was limited to the internal carotid artery proximal to the origin of the posterior communicating artery. In this study, coils were not allowed but 98% of subjects received more than one PED.

All these studies reported high incidences of successful FD deployment. Successful PED deployment was achieved in 97.9% in PITA and in 99% in PUFS. SFD failure rates were 4% in the SFD Registry and 1.5% in the French study; however, both studies described difficulties in opening the device and suboptimal positioning (17% and 12.3% respectively). These TAE are subjective and their described frequency probably depends on an assessment being mandated on case report forms (CRFs) in study protocols. Any consequent CAEs are likely to be more consistently measured. They were surprisingly infrequently reported in the SFD Registry in which only 1 patient experienced a new stroke despite a high rate of TAEs with inadvertent parent artery occlusion reported after 7 (10%) procedures. In the French study, parent artery occlusions occurred in 6 (9%) procedures but resulted in 4 strokes and one further patient suffered a stroke in the early post procedure period after FD thrombosis (See Table 1). It is probable that this complication is usually caused by poor conformity of the FD with the parent artery wall causing local thrombus and / thromboembolism. Stent conformity can be improved by stentplasty after deployment.

Table 1 Summary of published series (n>10) reporting clinical experiences after flow diverter treatments

In PITA, two patients (6.4%) suffered a major peri-procedural stroke. One was related to rupture of the target artery during balloon angioplasty that became necessary due to inappropriate PED deployment. In the second case an ischaemic stroke was attributed to potential occlusion of lenticulostriate arteries covered by multiple PEDs and a Neuroform stent [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed]. In PUFS the rate of major ipsilateral stroke and death at 180 days was 5.6%. Two of these events occurred peri-procedurally (1.9%); one was ischaemic due to PED thrombosis and the other haemorrhage following administration of a IIb-IIIa inhibitor for suspected minor ischaemic lesions [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads]. Szikora et al. reported a major subarachnoid haemorrhage resulting in death, a few hours post procedure that was attributed to rupture of a concurrent small aneurysm located on the same internal carotid artery but distal to the target lesion [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47. No peri-procedural CAEs were reported in the Buenos Aires experience [80]80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42. In a voluntary survey conducted by the Swiss Society of Minimally Invasive Neurological Therapy (ESMINT), peri-procedural thromboembolic events were reported in 6.89% of 697 patients [102]102. Szikora I, Kulcsar Z; Incidence of hemorrhagic complications following flow diverter treatment of intracranial aneurysms; Presented at the 3d ESMINT Congress, September 8-10, 2011; Nice, France.  Peri-procedural parenchymal haemorrhages were found in 1.0% and SAH in 0.86%, totalling an 8.75% procedural complication rate.

The immediate effect of FD deployment on aneurysm filling is described in most reports. Szikora et al. [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47 proposed an assessment scheme, which distinguishes complete stasis of contrast media within the sac from flow reduction. Rates for the former were 10% in the SFD Registry and 6.6% in the French study. Thus complete cessation of endosaccular blood flow is not the usual immediate response to FD deployment. Though symptoms generally improve, delayed symptomatic deteriorations may occur. Szikora et al. [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47 reported improvement in 68% of patients presented with compression syndromes or headaches and PUFS reported improvement of aneurysm related symptoms in 41% of patients [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads].

Back to top

Delayed Clinical Outcomes

a) Ischaemic and Compression Symptoms

The cause of delayed non-haemorrhagic CAEs is often the subject of speculation because it is sometimes difficult to distinguish ischaemic events due to reductions in parent artery blood flow or small branch artery occlusion from symptoms due to induced aneurysm thrombosis. The latter was reported in early FD reports [80]80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42 [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central] and occurred in the first post-treatment 2 weeks in 24/64 (37.5%) of the French study patients [73]73. Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al.; Flow-Diverter Silk Stent for the Treatment of Intracranial Aneurysms: 1-year Follow-Up in a Multicenter Study; Am J Neuroradiol. Epub 2012 Feb 2. In some cases peri-aneurysmal vasogenic oedema is evident on MRI with gadolinium enhancement around the aneurysm along with evolving signal in the endosaccular thrombus.  Berge et al. [103]103. Berge J, Tourdias T, Moreau J-F, Barreau X, Dousset V; Perianeurysmal Brain Inflammation after Flow-Diversion Treatment; Am J Neuroradiol 2011;32:1930-1934 [PubMed] followed 17 patients treated with FDs and correlated clinical outcomes with such MR imaging findings. Seven patients exhibited emergence of new neurological symptoms or worsening of previous symptoms and MRI showed signs of a peri-aneurysmal brain inflammatory reaction, which was absent for those without clinical deterioration. The symptomatic group had larger aneurysms embedded in brain parenchyma and at least 50% sac thrombosis when new or increased symptoms occurred. These symptoms resolved within 30 days in all but one patient and the abnormal signal on MRI had disappeared at 3 months. The authors concluded that induced thrombosis initiates an inflammatory reaction that spreads to the adjacent parenchyma and may be responsible for post-treatment exacerbation of symptoms. This reaction appears to be worse in larger aneurysms due to a greater thrombus load. It has parallels with the post-implantation syndrome observed after endovascular abdominal aortic aneurysm repair, and this similarly is the basis for treating patients with corticosteroids. Aneurysm swelling alone may cause or exacerbate compression symptoms and has been observed in thrombosed aneurysms after parent artery occlusion due to swelling of endosaccular thrombus [104]104. Byrne JV, Chung A; Volume changes in large and giant aneurysms demonstrated on MRI following Endovascular treatment; Neuroradiology 2000;43(5):415. Very late increases in the size of treated aneurysms have also been reported [85]85. Hampton T, Walsh D, Tolias C, Fiorella D; Mural destabilization after aneurysm treatment with a flow-diverting device: A report of two cases; J NeuroInterv Surgery 2011;3(2):167-171 after FD treatments and have raised concerns that endosaccular thrombosis once induced by FDs without ECE may not be stable over time.

Other causes of delayed ischaemic CAEs are late FD thrombosis, thromboembolism, transient ischaemic events (TIA) and narrowing of the vessel lumen (i.e. in-stent stenosis). Post-treatment symptomatic worsening in the Silk Registry (n=70) was attributed to in-stent thrombosis in 1, thromboembolism in 1 and aneurysm swelling or FD-covered branch artery occlusion in 4 patients [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central]. These symptoms resolved completely in half of the patients.  In PITA no delayed CAEs were recorded [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed]. In PUFS, 4 of 6 major ipsilateral stroke events occurred later than the first postoperative day. Two were ischaemic, one was related to significant in-stent stenosis and one was attributed to non-compliance with the antiplatelet protocol. Temporary aggravation of compression related symptoms was not reported but one patient (0.9%) suffered a partial visual field loss attributed to retinal ischaemia. Clinically, 41% of symptomatic patients improved within 180 days [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads]. In a study of 54 patients with 57 aneurysms treated with PED, McAuliffe et al. found no permanent morbidity and mortality despite asymptomatic in-stent stenosis at 6 months in two patients [105]105. W. McAuliffea, V. Wycocoa, H. Riceb, C. Phatourosa, T.J. Singha, J. Wenderoth; Immediate and Midterm Results following Treatment of Unruptured Intracranial Aneurysms with the Pipeline Embolization Device; AJNR American Journal of Neuroradiology 2012; 33: 164-170 [PubMed]. In another case series of 96 aneurysms and 5 dissections, Fisher et al. reported 6 (5.9%) major complications; 2 of which were attributed to FD thrombosis [106]106. Fischer S, Vajda Z, Aguilar Perez M, Schmid E, Hopf N, Bäzner H, Henkes H; Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections; Neuroradiology epub 2011 Sept 02; DOI: 10.1007/s00234-011-0948-x.

Post-treatment patency of perforators and sidewall branches arising from the vessel segment covered with high mesh density FDs remains a major theoretical concern despite the evidence from animal models that side branches covered by FDs remain patent, as described above [26, 27]26. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ; A new endoluminal, flow-disrupting device for treatment of saccular aneurysms; Stroke. 2007;38:2346 -235227. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ; A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms; AJNR Am J Neuroradiol. 2009;30:1153-1158 [PubMed]. Kulcsar et al. [94]94. Kulcsár Z, Ernemann U, Wetzel SG, et al.; High-profile flow diverter (Silk) implantation in the basilar artery: efficacy in the treatment of aneurysms and the role of the perforators; Stroke 2010;41:1690-96 addressed the issue in a small case series (n 12) of basilar artery aneurysms treated with FDs. They deployed SFD to cover one-third or more of the basilar artery length and subsequent angiograms confirmed patency of angiographically visible vessels covered by the FD in all but 1 patient. In this case, the proximal posterior cerebral artery failed to opacify. The patient was asymptomatic because of adequate collateral blood flow from the posterior communicating artery. In the Budapest experience, of 28 visible side branches covered by PEDs, 1 ophthalmic artery occluded intra-procedurally and 2 in a delayed fashion. These delayed occluded arteries were covered by 3 or 4 PEDs and both were asymptomatic. It is hypothesised that side branches with good collateral supply, such as the ophthalmic artery, have a higher tendency for occlusion due to pressure gradient changes following FD coverage [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47. The issue needs further study, particularly if the device is, in future, to be used in distal vessels with side branches whose diameters approach those of FD pore sizes.

It is probable that delayed asymptomatic branch artery narrowing or occlusion is underreported. Slow narrowing of arterial origins allows time for collateral blood flow to develop and is thus without clinical sequelae. It may be a ‘physiological’ response to a chronic flow reduction sufficient to induce collaterals. This effect should be distinguished from thrombosis of the FD, which may be a delayed and sudden event. Fiorella et al. [82]82. Fiorella D, Hsu D, Woo HH, Tarr RW, Nelson PK; Very late thrombosis of a pipeline embolization device construct: case report; Neurosurgery 2010; 67: ons E313-ons E314 reported stroke due to PED thrombosis at 23 months, and Klisch et al. [83]83. Klisch J, Turk A, Turner R, Woo HH, Fiorella D; Very late thrombosis of flow-diverting constructs after the treatment of large fusiform posterior circulation aneurysms; Am J Neuroradiology 2011; 32(4):627-32 12 months after treatments. Klisch et al. reported a patient who developed a fatal infarct and another who suffered no deficit after stopping APDs, illustrating how FD thrombosis may be asymptomatic and missed.

Comparing reports is complicated by variations in protocols for APD prophylaxis against stent thrombosis. There is general agreement among FD users of its value following treatment but peri-procedural prescription protocols vary between centres and studies. In the PITA trial [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed] 75mg clopidogrel and 100mg aspirin was required starting 48hrs prior to and continued for 6 weeks following procedures. Aspirin was given indefinitely afterwards. In PUFS, 325mg of aspirin was given for 48hrs and 75mg of clopidogrel for 7 days prior to procedures, but a loading dose of 600mg clopidogrel was allowed the day before procedure. Clopidogrel was required for 3 months, and aspirin for 6 months post procedure, but according to reports most (but not all) centres also used clopidogrel for 6 months [102]102. Szikora I, Kulcsar Z; Incidence of hemorrhagic complications following flow diverter treatment of intracranial aneurysms; Presented at the 3d ESMINT Congress, September 8-10, 2011; Nice, France. Platelet activity tests were not mandatory in any of these studies. The ESMINT FD related bleeding survey reported delayed thromboembolic events in 4.73% [102]102. Szikora I, Kulcsar Z; Incidence of hemorrhagic complications following flow diverter treatment of intracranial aneurysms; Presented at the 3d ESMINT Congress, September 8-10, 2011; Nice, France and in some reports, it was unclear whether these patients were adequately anti-aggregated. The effects of APDs on the time-scale of endothelialisation of FDs and whether long fusiform aneurysms take longer to ‘heal’ and so require longer APD treatment than saccular aneurysms is unknown. So far, no systematic study has addressed this issue.

b) Early and Delayed Intracranial Haemorrhage

Aneurysm bleeding is easily distinguished as a cause of delayed CAEs but its pathophysiology remains uncertain. Both early and delayed post-procedure aneurysmal ruptures have been reported. There have been several small series reports of spontaneous intracranial haemorrhage in the peri-procedural period (i.e. up to 7 days) [81]81. Cruz J P, Chow M, O'Kelly C, Marotta B, Spears J, Montanera W, Fiorella D, Marotta T; Delayed Ipsilateral Parenchymal Haemorrhage following Flow Diversion for the Treatment of Anterior Circulation Aneurysms; AJNR 2012; 33:603-8 [88]88. Phillips T, Mitchell P, Dowling R, Yan B; Endovascular treatment of intracranial aneurysms with new generation flow diverting stents. Early experience in an Australian neurointerventional centre; Proceedings of the 61st annual scientific meeting of the Royal Australian and New Zealand College of Radiologists (RANZCR) 2010; Oct 14-17; Perth, Australia Journal of Medical Imaging Radiation & Oncology 2010 Oct; 54(Supplement1):A122 [89]89. Hartmann M, Rohde S, Braun C, Hahnel S, Bendszus M; Endovascular treatment of cerebral aneurysms with the pipeline embolization device; Proceedings of the Jahrestagung der Deutschen Gesellschaft fur Neuroradiologie (DGNR) 2010; Sept 22-25; Mannheim Germany. Clinical Neuroradiology 2010;20(3):190-1. These have occurred due to aneurysm or parent artery ruptures or in the brain upstream to FD treatments. Cruz et al. [81]81. Cruz J P, Chow M, O'Kelly C, Marotta B, Spears J, Montanera W, Fiorella D, Marotta T; Delayed Ipsilateral Parenchymal Haemorrhage following Flow Diversion for the Treatment of Anterior Circulation Aneurysms; AJNR 2012; 33:603-8, reported parenchymal haemorrhage remote from PED treated aneurysms between 1 – 7 days in 4/47 patients which they were unable to explain on the basis of procedural perforations. Spontaneous haemorrhage may be exacerbated by the aggressive use of APDs. Extra-cranial bleeding has also been reported as a cause of substantial morbidity and this clearly can be attributed to APD use. In the SFD Registry [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central] 3 patients suffered complications due to acute extra-cranial bleeding and this contributed to one death.

The ESMINT survey [102]102. Szikora I, Kulcsar Z; Incidence of hemorrhagic complications following flow diverter treatment of intracranial aneurysms; Presented at the 3d ESMINT Congress, September 8-10, 2011; Nice, France was conducted in 2011 to assess the frequency of FD related bleeding complications. It collected data on 720 aneurysms in 697 patients and investigated the potential reasons for post-FD haemorrhagic complications. Bleeding was classified as intra-cerebral (parenchymal) haemorrhage (ICH) remote from the target aneurysms and SAH related to rupture of the target aneurysm. Delayed ICH was reported in 2.0% and late aneurysm rupture in 1.0% of patients. The reason for remote ICH is unknown. It has been speculated that thrombosis of a giant aneurysm may eliminate the damping effect of the aneurysm sac and subsequently increase the pulsatility of the distal vasculature resulting in rupture of distal small arterioles. Another theory is that ICH is due to repeated small emboli released from the surface of the yet non-endothelialised FD causing small, clinically silent infarcts, which undergo haemorrhagic transformation exacerbated by APD treatment. In this case, the ICH is primarily related to insufficient anti-aggregation.

There have been several reports of delayed rupture of treated aneurysms. Turowski et al. [98]98. Turowski B, Macht S, Kulcsar Z, et al.; Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (Silk stent): do we need to rethink our concepts? Neuroradiology 2011;53:37- 41. reported fatal haemorrhage after SFD treatment of a symptomatic, but unruptured, large para-ophthalmic carotid aneurysm which appeared substantially occluded on an initial CT.  Acute SAH occurred 20 days post-procedure and a second CT showed recannulation. Late bleeding from treated but incompletely occluded aneurysms has been reported by others [92, 62, 88, 76]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central]62. Mocco J, Snyder KV, Albuquerque FC, Bendok BR, Alan SB, Carpenter JS, et al.; Treatment of intracranial aneurysms with the Enterprise stent: a multicenter registry; J Neurosurg 2009;110:35-988. Phillips T, Mitchell P, Dowling R, Yan B; Endovascular treatment of intracranial aneurysms with new generation flow diverting stents. Early experience in an Australian neurointerventional centre; Proceedings of the 61st annual scientific meeting of the Royal Australian and New Zealand College of Radiologists (RANZCR) 2010; Oct 14-17; Perth, Australia Journal of Medical Imaging Radiation & Oncology 2010 Oct; 54(Supplement1):A12276. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed] and a similar, but much later, reopening of a thrombosed aneurysm was documented in the Silk Registry [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central]. In the ESMINT survey, all of the aneurysms which ruptured after FD treatments were larger than 10 mm, and all but one was larger than 18mm, suggesting that this complication is specific to large and giant aneurysms. The delayed aneurysm rupture rate was 2.15% among 511 aneurysms larger than 10mm [102]102. Szikora I, Kulcsar Z; Incidence of hemorrhagic complications following flow diverter treatment of intracranial aneurysms; Presented at the 3d ESMINT Congress, September 8-10, 2011; Nice, France.

To explore the subject of delayed aneurysmal haemorrhage further and to identify putative factors Kulcsar et al. [107]107. Kulcsár Z, Houdart E, Bonafe A, et al.; Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow diversion treatment; Am J Neuroradiol 2011;32:20-25 [PubMed] collected 13 cases of post-procedural aneurysmal rupture from 12 centres worldwide and found: (a) 11 of the 13 patients were symptomatic prior to treatment, (b) most patients were hypertensive, (c) all aneurysms were >15mm in size (mean 22±6mm) and (d) all saccular aneurysms had aspect ratio >1.6.  Though FDs caused contrast stagnation in all the aneurysms, a persistent in-flow jet was observed in 11 aneurysms. The authors considered this feature to be a contributory factor towards delayed aneurysm rupture. The effects of blood flow dynamics on the risks of late haemorrhage have also been studied by Cebral et al. who used computational fluid dynamics to show that endosaccular pressure increased after FD treatments in aneurysms that ruptured in the peri-operative period (up to 7 days) [108]108. Cebral JR, Mut F, Rasch M, Ceratto R, Lylyk P, Putman CM; Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment; Am J Neuroradiol 2011; 32;27-33.

Late delayed rupture has raised concern over the stability of endosaccular thrombus after FD treatments. At follow up imaging, all aneurysms in Kulcsar’s study [107]107. Kulcsár Z, Houdart E, Bonafe A, et al.; Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow diversion treatment; Am J Neuroradiol 2011;32:20-25 [PubMed] except one (for which thrombosis could not be evaluated) were occluded to a significant degree with thrombus (complete occlusion in 2, >90% occlusion in 5, and 50 – 90% occlusion in 5). Post-mortem histopathological examination in 2 patients showed intraluminal thrombi lacking significant organisation or cellular colonisation. Thrombus is known to undergo spontaneous lysis and renewal, and this may have been maintained by persistent low-grade intra-aneurysmal flow through the FD. These unstable thrombi are presumably the result of enzymes (e.g. proteases and elastases) generated within thrombus, which could cause thinning and weakening of the aneurysmal wall. Because aneurysm bleeding is rare after ECE, in which thrombus is generally stable with cicatrisation and reverse remodelling of the vessel wall, it has been suggested that larger aneurysms be treated with adjuvant endosaccular coils to increase the flow diverting effect of FDs and improve the stability of induced thrombus.

Back to top

Angiographic Outcomes

Follow up periods and imaging methods vary considerably between the published studies of FD treatments. Catheter angiography is generally considered the best method of demonstrating the extent of aneurysm and parent artery patency because of the obscuring effect of the FD. CTA and MRA can show vessel patency but are more useful at showing changes in aneurysm size. Large and giant aneurysms should be considered as stable and completely occluded only after they have lost the initial increase in T1 and T2 signal intensities associated with recent thrombosis and unorganised intra-saccular thrombus.  As with APD regimes, no consensus has emerged on imaging follow up protocols and various grading schemes have been described which recognise that residual aneurysm filling is common in the medium-term [69, 90]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-4790. O'Kelly C, Spears J, Chow M, Wong J, Silvaggio J, Boulton M, et al.; Canadian experience with the pipeline embolization device for repair of unruptured intracranial aneurysms; Canadian Journal of Neurological Sciences Conference: 46th Annual Congress of the Canadian Neurological Sciences Federation, 2011; 38(3) supp 1:S31 or are designed for assessing fusiform shape aneurysms [68]68. Kamran M, Yarnold J, Grunwald IQ, Byrne JV; Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema; Neuroradiology 2011; 53: 501-8 [PubMed].

Timing is important and angiographic results appear to be directly related to the interval since treatment and inversely to the use of APDs. In the SFD Registry [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central], no follow up protocol was pre-specified and imaging intervals varied between centres. Imaging was performed at various times up to 6 months after treatment in 49 of 70 treated aneurysms and showed complete occlusion in 49%, neck remnants in 26% and saccular filling in 24%.  In the PITA study, which specified assessments at 6 months, the rate for complete occlusion was 93.3% [76]76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed].  In PUFS, 73% of all study subjects and 80% of those having an angiogram at 6 months had complete occlusion [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads]. The rate of complete occlusion increased to 86% among those examined at 1 year. The lowest rate for successful occlusion reported in a substantial series (n = 96) is 69% by O’Kelly et al. [90]90. O'Kelly C, Spears J, Chow M, Wong J, Silvaggio J, Boulton M, et al.; Canadian experience with the pipeline embolization device for repair of unruptured intracranial aneurysms; Canadian Journal of Neurological Sciences Conference: 46th Annual Congress of the Canadian Neurological Sciences Federation, 2011; 38(3) supp 1:S31. In the single centre report by Lylyk et al. [80]80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42, complete occlusion rates were 56% at 3 months and 95% at 12 months. Similarly, Fisher et al reported 52% complete occlusion at 3 months and 74% at 10 months out of 101 aneurysms and arterial dissections [106]106. Fischer S, Vajda Z, Aguilar Perez M, Schmid E, Hopf N, Bäzner H, Henkes H; Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections; Neuroradiology epub 2011 Sept 02; DOI: 10.1007/s00234-011-0948-x and McAuliffe [105]105. W. McAuliffea, V. Wycocoa, H. Riceb, C. Phatourosa, T.J. Singha, J. Wenderoth; Immediate and Midterm Results following Treatment of Unruptured Intracranial Aneurysms with the Pipeline Embolization Device; AJNR American Journal of Neuroradiology 2012; 33: 164-170 [PubMed] complete occlusion in 62% at 1 month and 86% at 6 months.

Changes in the size of aneurysms after FD-induced thrombosis was studied using CT or MR by Berge et al. [73]73. Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al.; Flow-Diverter Silk Stent for the Treatment of Intracranial Aneurysms: 1-year Follow-Up in a Multicenter Study; Am J Neuroradiol. Epub 2012 Feb 2. Like Lylyk et al. [80]80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42, they observed improving occlusion rates with time (68% complete occlusion at 6 months and 84% at 12 months) and that at 12 months 30% of the completely occluded aneurysm had disappeared, 56% were smaller, 11% were unchanged and 7% larger. Adjuvant endosaccular coils reduced the chance of the aneurysm size reducing; 96% of those treated without coils but only 26% of those treated with adjuvant coils disappeared or reduced in size. None of the reported series have separately reported angiographic outcome for saccular and fusiform aneurysm shapes which is surprising since the parameters generally used, e.g. neck remnant and body filling, are often quite different as pointed out be Kamran et al. [68]68. Kamran M, Yarnold J, Grunwald IQ, Byrne JV; Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema; Neuroradiology 2011; 53: 501-8 [PubMed] and outcomes poorer [92]92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central].

There have been no large studies with follow up data beyond 1 year and reports of very late aneurysm rupture have raised concern about the long-term stability of the induced thrombosis. Szikora et al. found no recurrence of 11 aneurysm studied by MR/MRA at 1.5 to 2 years [69]69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47. The PUFS trial is following patients for up to 5 years and the 180 days and 1 year angiographic results are available. Ninety seven percent of aneurysms that were completely occluded at 180 days remained completely occluded on angiograms performed at 1 year without major stenosis or occlusion of the parent artery [77]77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads]. These data suggest that angiographic aneurysm occlusion achieved by FD is at least as stable as after ECE but data from long-term studies are still needed to confirm that delayed re-opening is a rare event.

Back to top

Ongoing randomised clinical trials for flow diverter devices

So far systematic study of the use of FDs is limited to non-comparative cohort studies. Randomised clinical trials have been proposed to compare it with conventional endovascular aneurysm treatments in a rigorous fashion. One prospective randomised controlled trial of the SFD, called MARCO POLO (Multicentre randomised trial on selective endovascular aneurysm occlusion with coils versus parent vessel reconstruction using the SILK flow diverter) is underway. It has been designed to compare SFD treatment against ECE for efficacy (angiographic outcomes at 12 months post-procedure) and safety (technical problems and clinical complications in a 12 month period post-procedure) [109]109. "Efficacy Trial of Intracranial Aneurysm Treatment Using Two Different Endovascular Techniques (Marco Polo)". Accessed March 2012. Available at [Clinical Trials].

Similar trials of the PED are underway which will evaluate the technical and clinical performance of this device. These include: (a) FIAT trial (Flow diversion in Intracranial Aneurysm Treatment), comparing clinical outcomes between flow diversion and best standard treatment (conservative, coiling, stenting or parent artery occlusion methods) [110]110. "Flow Diversion in Intracranial Aneurysm Treatment (FIAT)". Accessed March 2012. Available at [Clinical Trials] and (b) COCOA trial (Complete Occlusion of COilable intracranial Aneurysms) comparing PED with ECE in small wide neck aneurysms [111]111. D. Fiorella, P. Lylyk, I. Szikora, et al.; 'Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysms'; Journal of NeuroInterventional Surgery 2009; vol. 1, pp. 56-65.. Clearly, these data are needed before a substantial shift to treat most aneurysms with FDs occurs. The experience of the last 5 years suggests that with improving designs and physician experience, such a shift is now a definite possibility.

Conflict of interest

JVB's department receives research support from Balt Extrusion, and JVB is a paid scientific Consultant advisor for Codman Neurovascular. IS’s Department of Neurointervention at the National Institute of Neurosciences has previously received research support from Covidien/ev3, and IS consults for Covidien/ev3 and Stryker Neurovascular.

References

  1. Ferns SP, Sprengers MES, van Rooij WJ, Rinkel GJE, van Rijn JC, Bipat S, et al.; Coiling of Intracranial Aneurysms / Supplemental Appendix; Stroke 2009; 40:523-9
  2. Campi A, Ramzi N, Molyneux AJ, Summers PE, Kerr RS, Sneade M, Yarnold JA, Rischmiller J, Byrne JV; Retreatment of ruptured cerebral aneurysms in patients randomized by coiling or clipping in the International Subarachnoid Aneurysm Trial (ISAT); Stroke 2007; 38:1538-1544 [PubMed]
  3. Hope JK, Byrne JV, Molyneux AJ; Factors influencing successful angiographic occlusion of aneurysms treated by coil embolization; American Journal of Neuroradiology 1999; 20:391-399 [PubMed]
  4. White PM, Lewis SC, Gholkar A, Nahser H, Cognard C, Forrester L, et al.; Hydrogel-coated coils versus bare platinum coils for the endovascular treatment of intracranial aneurysms (HELPS): a randomised controlled trial; Lancet 2011;377:1655-1662
  5. Coley S, Sneade M, Clarke A, Mehta Z, Kallmes D, Cekirge S, et al.; Cerecyte Coil Trial: Procedural safety and Clinical outcomes in patients with ruptured and Unruptured Intracranial Aneurysms; Am J Neuroradiol 2012; epub Dec 2011
  6. Moret J, Cognard C, Weill A, Castaings L, Rey A; The “Remodelling Technique” in the treatment of wide neck intracranial aneurysms. Angiographic results and clinical follow-up in 56 cases; Interv Neuroradiol 1997;3(1):21-35
  7. Molyneux AJ, Cekirge S, Saatci I, Gál G; Cerebral Aneurysm Multicenter Swiss Onyx (CAMEO) trial: results of a prospective observational study in 20 Swiss centers; Am J Neuroradiol 2004 Jan;25(1):39-51
  8. Piotin M, Pistocchi S, Bartolini B, Blanc R; Intracranial aneurysm coiling with PGLA-coated coils versus bare platinum coils: long-term anatomic follow-up; Neuroradiology 2012 Apr;54(4):345-8. 2011.
  9. Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, Moret J; Stent assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms; Stroke 2010;41:110-5 [PubMed]
  10. Lieber BB, Stancampiano AP, Wakhloo AK; Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity; Ann Biomed Eng 1997;25:460-469 [PubMed]
  11. Imbesi SG, Kerber CW; Analysis of slipstream flow in a wide-necked basilar artery aneurysm: evaluation of potential treatment regimens; Am J Neuroradiol 2001;22(4), 721-724
  12. Imbesi SG, Knox K, Kerber CW; Aneurysm flow dynamics: alterations of slipstream flow for neuroendovascular treatment with liquid embolic agents; Am J Neuroradiol 2003;24(10), 2044-2049 [PubMed]
  13. Yu SC, Zhao JB; A steady flow analysis on the stented and non-stented sidewall aneurysm models; Med Eng Phys 1999;21(3), 133-141 [PubMed]
  14. Baráth K, Cassot F, Rüfenacht DA, Fasel JH; Anatomically shaped internal carotid artery aneurysm in vitro model for flow analysis to evaluate stent effect; Am J Neuroradiol 2004; 25(10), 1750-1759 [PubMed]
  15. Lieber BB, Livescu V, Hopkins LN, Wakhloo AK; Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow; Ann Biomed Eng 2002; 30:768 -777
  16. Canton G, Levy DI, Lasheras JC; Hemodynamic changes due to stent placement in bifurcating intracranial aneurysms; J Neurosurg 2005;103, 146-155 [PubMed]
  17. Cantón G, Levy DI, Lasheras JC, Nelson PK; Flow changes caused by the sequential placement of stents across the neck of sidewall cerebral aneurysms; J Neurosurg 2005; 103(5), 891-902 [PubMed]
  18. Seong J, Wakhloo AK, Lieber BB; In vitro evaluation of flow diverters in an elastase-induced saccular aneurysm model in rabbit; J Biomech Eng 2007;129;863-872
  19. Trager AL, Sadasivan C, Seong J, Lieber BB; Correlation between angio- graphic and particle image velocimetry quantifications of flow diverters in an in vitro model of elastase-induced rabbit aneurysms; J Biomech Eng. 2009;131:034506
  20. Wakhloo AK, Schellhammer F, de Vries J, Haberstroh J, Schumacher M; Self-expanding and balloon-expandable stents in the treatment of carotid aneurysms: An experimental study in a canine model; AJNR Am J Neuro- radiol. 1994;15:493-502.
  21. Turjman F, Massoud TF, Ji C, Guglielmi G, Vinuela F, Robert J; Combined stent implantation and endosaccular coil placement for treatment of experimental wide-necked aneurysms: a feasibility study in swine; Am J Neuroradiol 1994;15:1087-1090 [PubMed]
  22. Szikora I, Guterman LR, Wells KM, Hopkins LN; Combined use of stents and coils to treat experimental wide necked carotid aneurysms: preliminary results; Am J Neuroradiol 1994; 15:1091-1102 [PubMed]
  23. Massoud TF, Turjman F, Ji C, Viñuela F, Guglielmi G, Gobin YP, Duckwiler GR; Endovascular treatment of fusiform aneurysms with stents and coils: technical feasibility in a swine model; Am J Neuroradiol 1995; 16: 1953-1963
  24. Geremia G, Haklin M, Brennecke L; Embolization of experimentally created aneurysms with intravascular stent devices; Am J Neuroradiol 1994;15:1223-1231 [PubMed]
  25. Wakhloo AK, Tio FO, Lieber BB, Schellhammer F, Graf M, Hopkins LN; Self-expanding Nitinol stents in canine vertebral arteries: Hemodynamics and tissue response; Am J Neuroradiol 1995;16:1043-1051 [PubMed]
  26. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ; A new endoluminal, flow-disrupting device for treatment of saccular aneurysms; Stroke. 2007;38:2346 -2352
  27. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ; A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms; AJNR Am J Neuroradiol. 2009;30:1153-1158 [PubMed]
  28. Tenjin H, Asakura F, Nakahara Y; Evaluation of intraaneurysmal blood velocity by time-density curve analysis and digital subtraction angiography; Am J Neuroradiol 1998;19:1303-1307 [PubMed]
  29. Kawanishi M, Nagasawa S, Ohta T, Kajimoto S, Kondoh S; Simulation study on therapeutic vertebral artery occlusion for VA-PICA giant aneurysm; Neurol Res 1994;16, 100-103 [PubMed]
  30. Kim M, Taulbee DB, Tremmel M, Meng H; Comparison of two stents in modifying cerebral aneurysm hemodynamics; Ann Biomed Eng. 2008;36: 726 -741
  31. Asakura F, Tenjin H, Sugawa N, Kimura S, Oki F; Evaluation of intraaneurysmal blood flow by digital subtraction angiography: blood flow change after coil embolization; Surg Neurol 2003; 59(4), 310-319.
  32. Rudin S, Wang Z, Kyprianou I, Hoffmann KR, Wu Y, Meng H, Guterman LR, Nemes B, Bednarek DR, Dmochowski J, Hopkins LN; Measurement of flow modification in phantom aneurysm model: comparison of coils and a longitudinally and axially asymmetric stent - initial findings; Radiology 2004;231(1), 272-276. [PubMed]
  33. Sadasivan C, Lieber BB, Gounis MJ, Lopes DK, Hopkins LN; Angiographic quantification of contrast medium washout from cerebral aneurysms after stent placement; Am J Neuroradiol 2002;23(7):1214-1221. [PubMed]
  34. Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB; An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model; Stroke 2009; 40(3):952-958 [PubMed]
  35. Liou TM, Li YC, Juan WC; Numerical and experimental studies on pulsatile flow in aneurysms arising laterally from a curved parent vessel at various angles; J Biomech 2007;40(6), 1268-75
  36. Appanaboyina S, Lohner R, Scrivano E, Miranda C, Lylyk P, Putman C, et al.; Computational modelling of blood flow in side arterial branches after stenting of cerebral aneurysm; Int J Comput Fluid Dynamics 2008;22:669-676
  37. Aurboonyawat T, Blanc R, Schmidt P, et al.; An in vitro study of Silk stent morphology; Neuroradiology 2011;53:659-67. [PubMed]
  38. Pero G, Denegri F, Valvassori L, Boccardi E, Scialfa G; Treatment of a middle cerebral artery giant aneurysm using a covered stent: Case report; J Neurosurg 2006;104(6):965-8 [PubMed]
  39. Mase M, Banno T, Yamada K, Katano H; Endovascular stent placement for multiple aneurysms of the extracranial internal carotid artery: Technical case report; Neurosurgery 1995;37:832-835. [PubMed]
  40. Hurst R, Haskal Z, Zager E; Endovascular stent treatment of cervical internal carotid artery aneurysms with parent vessel preservation; Surg Neurol 1998;50:313-317
  41. Horowitz M, Miller G, Meyer Y, Carstens G, Purdy P; Use of intravascular stents in the treatment of internal carotid extracranial vertebral artery pseudoaneurysms; Am J Neuroradiol 1996;17:693-6 [PubMed]
  42. Marks MP, Dake MD, Steinberg GK, Norbash AM, Lane B; Stent placement for arterial and venous cerebrovascular disease: Preliminary experience; Radiology 1994. [PubMed]
  43. Higashida RT, Smith W, Gress D, Urwin R, Dowd CF, Balousek PA, Halbach VV; Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery. Case report and review of the literature; J Neurosurg 1997;87(6):944-9 [PubMed]
  44. Brassel F, Rademaker J, Haupt C, Becker H; Intravascular stent placement for a fusiform aneurysm of the posterior cerebral artery: Case report; Eur Radiol 2001;11:1250-1253 [PubMed]
  45. Lanzino G, Wakhloo AK, Fessler RD, Hartney ML, Guterman LR, Hopkins LN; Efficacy and current limitations of intravascular stents for intracranial internal carotid, vertebral, and basilar artery aneurysms; J Neurosurg 1999;91(4):538-546 [PubMed]
  46. Mericle RA, Lanzino G, Wakhloo AK, Guterman LR, Hopkins LN; Stenting and secondary coiling of intracranial internal carotid artery aneurysm: Technical case report; Neurosurgery 1998;43:1229-1234 [PubMed]
  47. Lylyk P, Cohen JE, Ceratto R, Ferrario A, Miranda C; Combined endovascular treatment of dissecting vertebral artery aneurysms by using stents and coils; J Neurosurg 2001; 94:427-432 [PubMed]
  48. Zenteno M, Modenesi Freitas JM, Aburto-Murrieta Y, Koppe G, Machado E, Lee A; Balloon-expandable stenting with and without coiling for wide-neck and complex aneurysms; Surg Neurol 2006; 66(6), 603-610 [PubMed]
  49. Wakhloo AK, Lanzino G, Lieber BB, Hopkins LN; Stents for intracranial aneurysms: the beginning of a new endovascular era? Neurosurgery 1998;43:377-379 [PubMed]
  50. Yang X, Wu Z, Mu S, Li Y, Lv M; Endovascular treatment of giant and large intracranial aneurysms using the neuroform stent-assisted coil placement; Neurol Res 2008;30: 598-602 [PubMed]
  51. Piotin M, Blanc R, Spelle L, Mounayer C, Piantino R, Schmidt PJ, Moret J; Stent assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms; Stroke 2010;41:110-5 [PubMed]
  52. Lylyk P, Cohen JE, Ceratto R, Ferrario A, Miranda C; Endovascular reconstruction of intracranial arteries by stent placement and combined techniques; J Neurosurg 2002; 97:1306-1313 [PubMed]
  53. Szikora I, Berentei Z, Kulcsar Z, Barath K, Berez A, Bose A, Nyary I; Endovascular treatment of intracranial aneurysms with parent vessel reconstruction using balloon and self expandable stents; Acta Neurochir 2006; 148:711-723
  54. Ahn JY, Han IB, Kim TG, Yoon PH, Lee YJ, Lee BH, Seo SH, Kim DI, Hong CK, Joo JY; Endovascular treatment of intracranial vertebral artery dissections with stent placement or stent-assisted coiling; Am J Neuroradiol 2006;27(7):1514-2
  55. Zenteno MA, Santos-Franco JA, Freitas-Modenesi JM, Gómez C, Murillo-Bonilla L, Aburto-Murrieta Y, Díaz-Romero R, Nathal E, Gómez-Llata S, Lee A; Use of the sole stenting technique for the management of aneurysms in the posterior circulation in a prospective series of 20 patients; J Neurosurg. 2008;108(6):1104-18
  56. Benndorf G, Campi A, Schneider GH, Wellnhofer E, Unterberg A; Overlapping stents for treatment of a dissecting carotid artery aneurysm; J Endovasc Ther 2001;8(6):566-70 [PubMed]
  57. Benndorf G, Herbon U, Sollmann WP, Campi A; Treatment of a ruptured dissecting vertebral artery aneurysm with double stent placement: Case report; Am J Neuroradiol 2001; 22: 1844-1848 [PubMed]
  58. Islak C, Kocer N, Albayram S; Bare stent-graft technique: a method of endoluminal vascular reconstruction for the treatment of giant and fusiform aneurysms; Am J Neuroradiol 2002;23:1589-1595 [PubMed]
  59. Doerfler A, Wanke I, Egelhof T, Stolke D, Forsting M; Double-stent method: therapeutic alternative for small wide-necked aneurysms. Technical note; JNeurosurg 2004; 100(1):150-4 [PubMed]
  60. Fiorella D, Albuquerque FT, Woo H, Rasmussen PA, Masaryk TJ, McDougall CG; Neuroform stent assisted aneurysm treatment: evolving treatment strategies, complications and results of long term follow-up; J NeuroIntervent Surg 2010; 2:16-22
  61. Lylyk P, Ferrario A, Pasbón B, Miranda C, Doroszuk G; Buenos Aires experience with the Neuroform self-expanding stent for the treatment of intracranial aneurysms; J Neurosurg 2005; 102(2):235-41
  62. Mocco J, Snyder KV, Albuquerque FC, Bendok BR, Alan SB, Carpenter JS, et al.; Treatment of intracranial aneurysms with the Enterprise stent: a multicenter registry; J Neurosurg 2009;110:35-9
  63. Sedat J, Chau Y, Mondot L, Vargas J, Szapiro J, Lonjon M; Endovascular occlusion of intracranial wide-necked aneurysms with stenting (Neuroform) and coiling: mid-term and long-term results; Neuroradiology 2009;51:401-9 [PubMed]
  64. Fiorella D, Albuquerque FC, Deshmukh VR, McDougall CG; Usefulness of the Neuroform stent for the treatment of cerebral aneurysms: results at initial (3-6-mo) follow-up; Neurosurgery 2005; 56:1191-201
  65. Biondi A, Janardhan V, Katz JM, Salvaggio K, Riina HA, Gobin YP; Neuroform stent-assisted coil embolization of wide-neck intracranial aneurysms: strategies in stent deployment and midterm follow-up; Neurosurgery 2007;61(3):460-8; discussion 468-9
  66. White PM, Lewis SC, Nahser H, Sellar RJ, Goddard T, Gholkar A; Hydrocoil Endovascular Aneurysm Occlusion and Packing Study (HELPS Trial): Procedural Safety and Operator-Assessed Efficacy Results; Am J Neuroradiol 2008;29(2):217-23. [PubMed]
  67. D’Urso PI, Lanzino G, Cloft HJ, Kallmes DF; Flow Diversion for Intracranial Aneurysms; Stroke 2011; 42: 2363-8
  68. Kamran M, Yarnold J, Grunwald IQ, Byrne JV; Assessment of angiographic outcomes after flow diversion treatment of intracranial aneurysms: a new grading schema; Neuroradiology 2011; 53: 501-8 [PubMed]
  69. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, et al.; Treatment of intracranialaneurysms by functional reconstruction of the parent artery: the Budapest experience with Pipeline embolization device; Am J of Neuroradiology 2010;31(6):1139-47
  70. Lubicz B, Collignon L, Raphaeli G, Pruvo J-P, Bruneau M, De Witte O, et al.; Flow-Diverter Stent for the Endovascular Treatment of Intracranial Aneurysms; Stroke 2010; 41: 2247-53 [PubMed]
  71. Byrne JV; Tutorials in Endovascular Neurosurgery and Interventional Neuroradiology; Springer 2012; Tutorial 6: 97-107
  72. Mangiacapra F, Muller O, Ntalianis A, Trana C, et al.; Comparison of 600 versus 300-mg Clopidogrel loading dose in patients with ST-segment elevation myocardial infarction undergoing primary coronary angioplasty; Am J Cardiol. 2010;106:1208-11 [PubMed]
  73. Berge J, Biondi A, Machi P, Brunel H, Pierot L, Gabrillargues J, et al.; Flow-Diverter Silk Stent for the Treatment of Intracranial Aneurysms: 1-year Follow-Up in a Multicenter Study; Am J Neuroradiol. Epub 2012 Feb 2
  74. Matouk C, O'Kelly C, Blis M, Sarma D, Spears J, et al.; Pipeline embolisation device reconstruction of ruptured intracranial aneurysms: Report of 2 cases; Proceedings of the 45th. Annual Congress of the Canadian Neurological Sciences Federation 2010 Quebec. Canadian Journal of Neurological Sciences 2010;3 (supplement 1): 888-889
  75. McAuliffe W, Wenderoth JD; Immediate and midterm results following treatment of recently ruptured intracranial aneurysms with the Pipeline embolization device; AJNR Am J Neuroradiol. 2012 Mar;33(3):487-93
  76. Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D; The pipeline embolization device for the intracranial treatment of aneurysms trial; Am J Neuroradiol 2011;32(1):34-40 [PubMed]
  77. Food & Drugs Administration (FDA). Chestnut Medical Technologies. Pipeline Embolization Device Executive Summary P100018 [report online]. 2011 Feb 1. Accessed August 31st 2011. Available at [FDA Downloads]
  78. Fiorella D, Woo HH, Albuquerque FC, Nelson PK; Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device; Neurosurgery 2008;62(5):1115-20.
  79. Fiorella D, Kelly ME, Albuquerque FC, Nelson PK; Curative reconstruction of a giant mid basilar trunk aneurysm with the pipeline embolization device; Neurosurgery 2009; 64:212-7
  80. Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, et al.; Curative endovascularreconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience; Neurosurgery 2009;64:632-42
  81. Cruz J P, Chow M, O'Kelly C, Marotta B, Spears J, Montanera W, Fiorella D, Marotta T; Delayed Ipsilateral Parenchymal Haemorrhage following Flow Diversion for the Treatment of Anterior Circulation Aneurysms; AJNR 2012; 33:603-8
  82. Fiorella D, Hsu D, Woo HH, Tarr RW, Nelson PK; Very late thrombosis of a pipeline embolization device construct: case report; Neurosurgery 2010; 67: ons E313-ons E314
  83. Klisch J, Turk A, Turner R, Woo HH, Fiorella D; Very late thrombosis of flow-diverting constructs after the treatment of large fusiform posterior circulation aneurysms; Am J Neuroradiology 2011; 32(4):627-32
  84. van Rooij WJ, Sluzewski M; Perforator infarction after placement of a pipeline flow-diverting stent for an unruptured A1 aneurysm; Am J Neuroradiol 2010;31:43-44
  85. Hampton T, Walsh D, Tolias C, Fiorella D; Mural destabilization after aneurysm treatment with a flow-diverting device: A report of two cases; J NeuroInterv Surgery 2011;3(2):167-171
  86. Fiorella D, Albuquerque F, Gonzalez F, McDougal CG, Nelson PK; Reconstruction of the right anterior circulation with the Pipeline embolization device to achieve treatment of a progressively symptomatic, large carotid aneurysm; J NeuroInterv Surgery 2009; 2:31-37
  87. Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Gubucz I, Nelson PK, et al.; Effect of flow modification on aneurysm induced mass effect; Proceedings of the 19th Symposium Neuroradiologicum - the World Congress of Diagnostic and Therapeutic Neuroradiology; Bologna, Italy. Neuroradiology 2010;23(1):324
  88. Phillips T, Mitchell P, Dowling R, Yan B; Endovascular treatment of intracranial aneurysms with new generation flow diverting stents. Early experience in an Australian neurointerventional centre; Proceedings of the 61st annual scientific meeting of the Royal Australian and New Zealand College of Radiologists (RANZCR) 2010; Oct 14-17; Perth, Australia Journal of Medical Imaging Radiation & Oncology 2010 Oct; 54(Supplement1):A122
  89. Hartmann M, Rohde S, Braun C, Hahnel S, Bendszus M; Endovascular treatment of cerebral aneurysms with the pipeline embolization device; Proceedings of the Jahrestagung der Deutschen Gesellschaft fur Neuroradiologie (DGNR) 2010; Sept 22-25; Mannheim Germany. Clinical Neuroradiology 2010;20(3):190-1
  90. O'Kelly C, Spears J, Chow M, Wong J, Silvaggio J, Boulton M, et al.; Canadian experience with the pipeline embolization device for repair of unruptured intracranial aneurysms; Canadian Journal of Neurological Sciences Conference: 46th Annual Congress of the Canadian Neurological Sciences Federation, 2011; 38(3) supp 1:S31
  91. Sararols L, Castillo L, Graell X, Macho J, San-Roman L, Macaya J, et al.; Right giant internal carotid artery bifurcation aneurism: Presentation with homonymous left hemianopsia and successful treatment with intraneurismatic bypass; Neuro-Ophthalmology Conference: 10th. Swiss Neuro-Ophthalmology Society, EUNOS Meeting Barcelona Spain. Conference Publication 2011; 35: S65
  92. Byrne JV, Beltechi R, Yarnold JA, Birks J, Kamran M; Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study; PLoS ONE 2010; 5: e12492. [PubMed] [PubMed Central]
  93. Lubicz B, Collignon L, Raphaeli G, et al.; Flow-diverter stent for the endovascular treatment of intracranial aneurysms: a prospective study in 29 patients with 34 aneurysms; Stroke 2010;41:2247-53 [PubMed]
  94. Kulcsár Z, Ernemann U, Wetzel SG, et al.; High-profile flow diverter (Silk) implantation in the basilar artery: efficacy in the treatment of aneurysms and the role of the perforators; Stroke 2010;41:1690-96
  95. Wagner A, Cortsen M, Hauerberg J, Romner B, Wagner MP; Treatment of intracranial aneurysms: Reconstruction of the parent artery with flow-diverting (Silk) stent; Neuroradiology 2011; Epub Sept 2011
  96. Cirillo L, Dalla'olio M, Princiotta C, Simonetti L, Stafa A, Toni F, Leonardi M; The use of flow-diverting stents in the treatment of giant cerebral aneurysms: Preliminary results; The Neuroradiology Journal 2010; 23: 220-224
  97. Kulcsár Z, Wetzel SG, Augsburger L, Gruber A, Wanke I, Rüfenacht DA; Effect of flow diversion treatment on very small ruptured aneurysms; Neurosurgery 2010; 67: 789-93 [PubMed]
  98. Turowski B, Macht S, Kulcsar Z, et al.; Early fatal hemorrhage after endovascular cerebral aneurysm treatment with a flow diverter (Silk stent): do we need to rethink our concepts? Neuroradiology 2011;53:37- 41.
  99. Pumar JM, Garcia-Dorrego R, Nieto A, Vazquez-Herrero F, Blanco-Ulla M, Vazquez-Martin A; Vascular reconstruction of a fusiform basilar aneurysm with the Silk embolization system; J Neurointerv Surg 2010; 2: 242-4 [PubMed]
  100. Leonardi M, Dall'olio M, Princiotta C, Simonetti L; Treatment of carotid siphon aneurysms with a microcell stent: A case report; Interv Neuroradiol 2008;14(4);429-434
  101. Rasskazoff S, Silvaggio J, Brouwer PA, Kaufmann A, Nistor A, Iancu D; Endovascular treatment of a ruptured blood blister-like aneurysm with a flow diverting stent; Interv Neuroradiol 2010;16(3):255-258. [PubMed] [PubMed Central]
  102. Szikora I, Kulcsar Z; Incidence of hemorrhagic complications following flow diverter treatment of intracranial aneurysms; Presented at the 3d ESMINT Congress, September 8-10, 2011; Nice, France
  103. Berge J, Tourdias T, Moreau J-F, Barreau X, Dousset V; Perianeurysmal Brain Inflammation after Flow-Diversion Treatment; Am J Neuroradiol 2011;32:1930-1934 [PubMed]
  104. Byrne JV, Chung A; Volume changes in large and giant aneurysms demonstrated on MRI following Endovascular treatment; Neuroradiology 2000;43(5):415
  105. W. McAuliffea, V. Wycocoa, H. Riceb, C. Phatourosa, T.J. Singha, J. Wenderoth; Immediate and Midterm Results following Treatment of Unruptured Intracranial Aneurysms with the Pipeline Embolization Device; AJNR American Journal of Neuroradiology 2012; 33: 164-170 [PubMed]
  106. Fischer S, Vajda Z, Aguilar Perez M, Schmid E, Hopf N, Bäzner H, Henkes H; Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections; Neuroradiology epub 2011 Sept 02; DOI: 10.1007/s00234-011-0948-x
  107. Kulcsár Z, Houdart E, Bonafe A, et al.; Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow diversion treatment; Am J Neuroradiol 2011;32:20-25 [PubMed]
  108. Cebral JR, Mut F, Rasch M, Ceratto R, Lylyk P, Putman CM; Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment; Am J Neuroradiol 2011; 32;27-33
  109. "Efficacy Trial of Intracranial Aneurysm Treatment Using Two Different Endovascular Techniques (Marco Polo)". Accessed March 2012. Available at [Clinical Trials]
  110. "Flow Diversion in Intracranial Aneurysm Treatment (FIAT)". Accessed March 2012. Available at [Clinical Trials]
  111. D. Fiorella, P. Lylyk, I. Szikora, et al.; 'Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysms'; Journal of NeuroInterventional Surgery 2009; vol. 1, pp. 56-65.